Instrument Science Report CAL/FOS-068 "FOS Red Detector Plate Scale and Orientation" November 1991 By: B. Bhattacharya and G. Hartig Abstract excerpted from the report: Revised plate scale factors, as determined from SV 1528 data, are presented. Analysis of these data indicates changes in plate scale of -0.4% +/- 0.53% along the diode array and 2.9% perpendicular to the array. The latter result is questionable due to the uncertainty in effective diode height. A negli- gible change in orientation of the x-y axis of 0.24 degrees +/- 0.32 degrees is found. These numbers indicate that the present scale factors of 1.43" per 256 y-base units and 0.357" per diode width are sufficiently accurate and do not need to be changed. For more information, please contact the above authors.