Laboratory Calibration of FOS Throughput George Hartig Space Tetescope Science Institute Instrument Science Report CAL/FOS-053 January 1989 Abstract Significant changes have been made to tbe FOS optical and detector systems since the previous absolute throughput measurements were reported. New calibration data were obtained, in ambient, in March and August, 1988, following these modifications. While the blue side efficiencies are slightly greater than those measured in 1984, marked improvements have been made in the red side throughput, as a result of detector replacement, optical realignment, and re-coating of the collimator mirror. However, FOS model calculations continue to overpredict the throughput of both sides of the spectrograph, in a wavelength independent manner.