! File: 1564C.PROP ! Database: PEPDB ! Date: 16-FEB-1994:23:23:10 coverpage: title_1: HRS FOCUS TEST proposal_for: OV/HRS pi_fname: JOHN pi_lname: TROELTZSCH pi_inst: BALL AEROSPACE SYSTEMS DIV. pi_country: USA pi_phone: 303-939-5781 keywords_1: ORBITAL VERIFICATION hours_par: 0.20 hrs: X pi_position: OV TEST ENGINEER ! end of coverpage abstract: line_1: This test defines the procedure used to evaluate HRS detector focus for both line_2: detectors. The detector mask edges are used to examine the focus for 5 line_3: values of the tweak current. This test is included in HRS Standard High line_4: Voltage Turn On, and does not need to be scheduled separately in the OV line_5: timeline. It will be used if problems are detected during the high voltage line_6: turn on test, or the detectors are found to be out of focus. The results of line_7: this test will be used to update the HRS high voltage and tweak values. ! ! end of abstract general_form_proposers: lname: KRUEGER fname: VERN inst: GODDARD SPACE FLIGHT CENTER country: USA ! lname: GARNER fname: HARRY inst: BALL AEROSPACE SYSTEMS DIV. country: USA ! lname: EBBETS fname: DENNIS inst: BALL AEROSPACE SYSTEMS DIV. country: USA ! lname: WILLIAMS fname: BRUCE inst: BALL AEROSPACE SYSTEMS DIV. country: USA ! lname: TROELTZSCH fname: JOHN inst: BALL AEROSPACE SYSTEMS DIV. country: USA ! ! end of general_form_proposers block general_form_text: question: 2 section: 1 line_1: This test will provide the results needed to optimize the values of the HRS line_2: high voltage and tweak current. ! question: 3 section: 1 line_1: A series of high resolution X scans are performed using 1 deflection step line_2: between each sample with the flat field lamp on using detector 1. The scans line_3: will step across the detector mask edge at center Y deflections. Each scan line_4: will be done with one of five tweak values chosen to bracket the default line_5: tweak value. The procedure is then repeated for side 2. ! question: 5 section: 1 line_1: During OV this test does not need to be scheduled, but should be available in line_2: case the results of the high voltage turn on indicate the detector is out of line_3: focus. It may also be used if routine science data indicates refocusing of the line_4: detector(s) is necessary. line_6: Prerequisites: line_8: 1. HRS High Voltage Turn On must have been completed and high voltage must be line_9: on. ! question: 7 section: 1 line_1: 1. The sharpness of mask edge will be analyzed to determine the ideal high line_2: voltage and tweak setting. ! question: 8 section: 1 line_1: This is an internal test and does not require any special pointing of the line_2: telescope. This test does not require the HRS to be the prime instrument. line_4: There is no target list with this proposal. ! question: 10 section: 1 line_1: 1. Test design support. line_3: 2. Test execution support team. line_5: 3. Data analysis facilities. line_7: 4. Data analysis personnel. ! !end of general form text general_form_address: lname: KRUEGER fname: VERN inst: GODDARD SPACE FLIGHT CENTER addr_1: GSFC CODE 683 city: GREENBELT state: MD zip: 20771 country: USA phone: 301-286-5004 ! lname: TROELTZSCH fname: JOHN category: PI inst: BALL AEROSPACE SYSTEMS DIV. addr_1: P.O.BOX 1062 city: BOULDER state: CO zip: 80306 country: USA phone: 303-939-4816 ! ! end of general_form_address records ! No fixed target records found ! No solar system records found ! No generic target records found exposure_logsheet: linenum: 1.000 sequence_1: DEFINE sequence_2: FOCUS targname: INTFLAT config: HRS opmode: ACCUM aperture: # num_exp: 1 time_per_exp: 231S priority: 1 param_1: DOPPLER=OFF, param_2: STEP-TIME=1.0 comment_1: USE DEFAULT OST ZLONGSCNX WITH comment_2: INITIAL DEFLECTIONS X=1622,Y=2048, comment_3: DELTA X=1. GENERATES 33 LINES OF DATA. ! linenum: 2.000 sequence_1: USE FOCUS aperture: SAFE1 req_1: SEQ 2-6 NO GAP; comment_1: SET DETECTOR 1 TWEAK VALUE TO OCTAL comment_2: 170. PLACE DATA FROM 2-6 ON TAPE comment_3: AT 32KBPS. ! linenum: 3.000 sequence_1: USE FOCUS aperture: SAFE1 comment_1: SET DETECTOR 1 TWEAK VALUE TO OCTAL comment_2: 174. ! linenum: 4.000 sequence_1: USE FOCUS aperture: SAFE1 comment_1: SET DETECTOR 1 TWEAK VALUE TO OCTAL comment_2: 204. ! linenum: 5.000 sequence_1: USE FOCUS aperture: SAFE1 comment_1: SET DETECTOR 1 TWEAK VALUE TO OCTAL comment_2: 210. ! linenum: 6.000 sequence_1: USE FOCUS aperture: SAFE1 comment_1: SET DETECTOR 1 TWEAK VALUE TO THE comment_2: SIDE 1 DEFAULT VALUE. ! linenum: 7.000 sequence_1: USE FOCUS aperture: SAFE2 req_1: SEQ 7-11 NO GAP; comment_1: SET DETECTOR 2 TWEAK VALUE TO OCTAL comment_2: 170. PLACE DATA FROM 7-11 ON TAPE comment_3: AT 32KBPS. ! linenum: 8.000 sequence_1: USE FOCUS aperture: SAFE2 comment_1: SET DETECTOR 2 TWEAK VALUE TO OCTAL comment_2: 174. ! linenum: 9.000 sequence_1: USE FOCUS aperture: SAFE2 comment_1: SET DETECTOR 2 TWEAK VALUE TO OCTAL comment_2: 204. ! linenum: 10.000 sequence_1: USE FOCUS aperture: SAFE2 comment_1: SET DETECTOR 2 TWEAK VALUE TO OCTAL comment_2: 210. ! linenum: 11.000 sequence_1: USE FOCUS aperture: SAFE2 comment_1: SET DETECTOR 2 TWEAK VALUE TO THE comment_2: SIDE 2 DEFAULT VALUE. ! ! end of exposure logsheet ! No scan data records found